20 January 2019 : Clinical Research
A Usability Study of 3 Radiotherapy Systems: A Comparative Evaluation Based on Expert Evaluation and User Experience
Mingyin Jiang12ABCDEF, Shenglin Liu12AEFG, Jiaqi Gao12BC, Qingmin Feng12BC, Qiang Zhang12ACDEFG*DOI: 10.12659/MSM.913160
Med Sci Monit 2019; 25:578-589
Abstract
BACKGROUND: The complex user interface design of radiotherapy treatment delivery systems can lead to use error and patient harm. In this study, we present the results of a comparison of 3 radiotherapy treatment delivery systems now used in China.
MATERIAL AND METHODS: We conducted a comprehensive usability study of 3 radiotherapy treatment delivery systems. Expert evaluation was performed through heuristic evaluation with 3 human-factors experts and 1 experienced radiation therapist for each system. User experience was assessed through perceived system usability and workload, using the National Aeronautics and Space Administration Task Load Index and the Post-Study System Usability Questionnaire.
RESULTS: For the expert evaluation, 47 usability problems were identified for Varian Trilogy, 75 for Elekta Precise, and 37 for Shinva XHA600E. Most problems were classified as major and minor usability problems, and were found in the process of patient setup and setup verification. For the user experience, radiation therapists presented a lower workload for Varian Trilogy compared to Elekta Precise (P<0.01) and Shinva XHA600E (P<0.01), and a lower workload for Elekta Precise compared to Shinva XHA600E (P=0.020). Radiation therapists perceived a higher system usability for Varian Trilogy compared to Shinva XHA600E (P<0.01), and a higher system usability for Elekta Precise compared to Shinva XHA600E (P<0.01).
CONCLUSIONS: This research provides valuable data on how 3 radiotherapy treatment delivery systems compare. The results of this study may be useful for hospital equipment procurement decisions, and designing next-generation products to improve patient safety.
Keywords: Equipment Design, Equipment Failure Analysis, Equipment Safety, Radiotherapy, User-Computer Interface, Patient safety, Patient Satisfaction, Surveys and Questionnaires
Editorial
01 July 2026 : Editorial
Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development AcceleratesDOI: 10.12659/MSM.954627
Med Sci Monit 2026; 32:e954627
In Press
Clinical Research
Prognostic Value of Mortality Scoring Systems in Patients With Severe Burns: Identifying Key Predictors of ...Med Sci Monit In Press; DOI: 10.12659/MSM.951713
Laboratory Research
Evaluation of the Trueness and Precision of Cast, Milled-Cast, Milled, and 3D-Printed Post-and-Core Techniq...Med Sci Monit In Press; DOI: 10.12659/MSM.953491
Clinical Research
Outcomes After Minimally Invasive Intramedullary Nail Fixation and Locking Plate Fixation Among Patients Wi...Med Sci Monit In Press; DOI: 10.12659/MSM.952670
Laboratory Research
Influence of Scan Body Geometry, Implant Angulation, and Interimplant Distance on the Accuracy of Maxillary...Med Sci Monit In Press; DOI: 10.12659/MSM.953734
Most Viewed Current Articles
17 Jan 2024 : Review article 14,176,260
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,758,016
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,466,166
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 708,825
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387