07 July 2018 : Laboratory Research  

Analysis of Base Monomer Elution from 3 Flowable Bulk-Fill Composite Resins Using High Performance Liquid Chromatography (HPLC)

Ryta Łagocka1ABDEF*, Małgorzata Mazurek-Mochol2EF, Katarzyna Jakubowska3AC, Maja Bendyk-Szeffer1F, Dariusz Chlubek3G, Jadwiga Buczkowska-Radlińska1G

DOI: 10.12659/MSM.907390

Med Sci Monit 2018; 24: LBR4679-4690

Abstract

BACKGROUND: The aim of this study was to evaluate the elution of BisGMA, UDMA, TEGDMA, and HEMA monomers from flowable bulk fill composite resins with different resin matrix compositions, polymerized in 4-mm-thick layers, into 3 elution media.

MATERIAL AND METHODS: Three bulk-fill (SDR® (SDR), X-tra base (XB) and BEAUTIFIL-Bulk Flowable (BF)) resin-composites were tested. Cylindrical samples were immersed in 100% ethanol, 75% ethanol, and distilled water. The concentrations of the monomers were measured using the HLPC method (Agilent Technologies 1200 Series) after 1 and 24 h, as well as after 3, 7, 14, and 21 days.

RESULTS: After polymerization of the tested resins, there was elution of the BisGMA, UDMA, TEGDMA, and HEMA monomers from the SDR and BF composites, but none of the tested monomers could be detected eluting from XB. The highest penetrations of the polymerized SDR and BF composites were observed in the 100% ethanol solution. This extraction medium eluted the highest amounts of free monomers. Some eluted monomers were not described in the composites Material Safety Data Sheets.

CONCLUSIONS: The elution of the residual monomers depended on the resin composition and the materials filler/resin matrix ratio. In composite materials, toxicity assessment should be carried out, and should consider both the material composition as given by the manufacturer, and also the residual monomers that elute from the polymerized material. The elution concentration and time of monomers from composites depended on the solvent used. The highest penetrations of the polymerized SDR and BF composites were observed in the 100% ethanol solution, and this extraction medium eluted the highest amounts of free monomers. The 75% ethanol was a more aggressive medium than water in terms of monomer elution from bulk fill composites.

Keywords: Chromatography, High Pressure Liquid, Composite Resins, Methylmethacrylate

Add Comment 0 Comments

Editorial

01 July 2026 : Editorial  

Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development Accelerates

Dinah V. Parums ORCID logo

DOI: 10.12659/MSM.954627

Med Sci Monit 2026; 32:e954627

0:00

In Press

Clinical Research  

Association Between Total Cholesterol–to–High-Density Lipoprotein Ratio and Gestational Hypertension: A Cas...

Med Sci Monit In Press; DOI: 10.12659/MSM.952395  

Review article  

Clinical Use of Endotracheal Intubation Without Neuromuscular Blockade: The Current Stage of Knowledge

Med Sci Monit In Press; DOI: 10.12659/MSM.951765  

Clinical Research  

Effect of Dexmedetomidine Hydrochloride Nasal Spray on Anxiety and Sleep in Patients Undergoing Gynecologic...

Med Sci Monit In Press; DOI: 10.12659/MSM.952465  

Clinical Research  

Prognostic Value of Mortality Scoring Systems in Patients With Severe Burns: Identifying Key Predictors of ...

Med Sci Monit In Press; DOI: 10.12659/MSM.951713  

Most Viewed Current Articles

17 Jan 2024 : Review article   14,176,173

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

13 Nov 2021 : Clinical Research   3,757,769

Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...

DOI :10.12659/MSM.932788

Med Sci Monit 2021; 27:e932788

0:00

14 Dec 2022 : Clinical Research   2,466,143

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research   708,797

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750