23 February 2018 : Laboratory Research
Finite Element Study of Implant Subsidence and Medial Tilt in Agility Ankle Replacement
Yunwei Cui1BCE, Pan Hu1BC, Ning Wei1CDE, Xiaodong Cheng1CDF, Wenli Chang1DF, Wei Chen1A*DOI: 10.12659/MSM.906151
Med Sci Monit 2018; 24: LBR1124-1131
Abstract
BACKGROUND: Clinical studies indicate that in total ankle arthroplasty, postoperative implant subsidence and medial tilt become two significant concerns of the ankle replacement system, and which are associated with the contact between the bones and the talar component. Up to now, little attention has focused on the contact between the bones and the talar component.
MATERIAL AND METHODS: In order to address implant subsidence and medial tilt, one three-dimensional finite element model of contact between the bone and the talar components was built with the material properties of the cancellous bone interpolated from the experimental data, which represents variation of material properties through the cancellous bones. The finite element model was used to study the following: variation of the Young’s modulus of the bones, stiffness of the talar component, loading direction, and loading magnitude with the implant subsidence.
RESULTS: The computational results reveal that a variety of Young’s modulus of the cancellous bones causes the medial tilting of the talar component and that big plastic strains are associated with tilting. The implant subsidence increases from 0.169 mm to 0.269 mm when the loading changes from 272 kg to 408 kg. However, to the contrary, the implant subsidence decreases from 0.2676 mm to 0.1783 mm when Young’s modulus of the bones increases 50%. However, the implant subsidence shows little change with a different Young’s modulus of the talar component from 88 GPa to 132 GPa.
CONCLUSIONS: Our study indicates that a variety of different Young’s modulus of the cancellous bones cause the medial tilting of the talar component. To solve subsidence and tilting, both the contact area and the variation of material properties should be taken into account.
Keywords: ankle fractures, Ankle Joint, Finite Element Analysis
Editorial
01 July 2026 : Editorial
Editorial: The WHO Identifies Ebola Disease Due to Bundibugyo Virus as a Public Health Emergency of International Concern (PHEIC) as Vaccine Development AcceleratesDOI: 10.12659/MSM.954627
Med Sci Monit 2026; 32:e954627
In Press
Review article
Clinical Use of Endotracheal Intubation Without Neuromuscular Blockade: The Current Stage of KnowledgeMed Sci Monit In Press; DOI: 10.12659/MSM.951765
Clinical Research
Effect of Dexmedetomidine Hydrochloride Nasal Spray on Anxiety and Sleep in Patients Undergoing Gynecologic...Med Sci Monit In Press; DOI: 10.12659/MSM.952465
Clinical Research
Prognostic Value of Mortality Scoring Systems in Patients With Severe Burns: Identifying Key Predictors of ...Med Sci Monit In Press; DOI: 10.12659/MSM.951713
Laboratory Research
Evaluation of the Trueness and Precision of Cast, Milled-Cast, Milled, and 3D-Printed Post-and-Core Techniq...Med Sci Monit In Press; DOI: 10.12659/MSM.953491
Most Viewed Current Articles
17 Jan 2024 : Review article 14,176,214
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
13 Nov 2021 : Clinical Research 3,757,839
Acceptance of COVID-19 Vaccination and Its Associated Factors Among Cancer Patients Attending the Oncology ...DOI :10.12659/MSM.932788
Med Sci Monit 2021; 27:e932788
14 Dec 2022 : Clinical Research 2,466,153
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research 708,809
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387