Suppression of power-line interference by analog notch filtering in the ECG signal for heart rate variability analysis: to do or not to do?
Laszlo HejjelMed Sci Monit 2004; 10(1): MT6-13 :: ID: 11556
Abstract
Background:Heart rate variability analysis has developed recently to a promising field of cardiology. Measures of heart rate variability can be corrupted by environmental noise or technical artifacts arising from analog and digital signal processinMaterial/Methods: The present investigation addressed the analysis of the effects of AC interference and its filtering on the precision and accuracy of heart rate detection. Artificial ECG recordings with predefined parameters were simulated by a computer and a data acquisition card, consecutively filtered by an analog notch filter (Q=2, 5 and 10) and digitized at 1 kHz by the same card ahead of and subsequent to the filter circuit. These latter signals were analyzed: the relative time points corresponding to the uncorrupted series at 7 points of the ventricular complexes were obtained and their means, standard deviations and maxima were computed separately.Results: The filtering of uncorrupted ECG signals does not result in heart rate period deviations greater than 1 ms. Power-line interference contamination proportionally alters the accuracy of representative point detection (maximal shift: 5% – 2 ms, 10% – 4 ms, 25% – 8 ms, 50% – 14 ms), which is consistently restored by notch filtering with a slight error (L1 ms on the ascending slope and at the peak). The descending slope is more distorted at filter Q=2 and at lower ventricular complex width; however this is constant, which also results in reliable RR-interval detection.Conclusions: AC notch filtering of ECG signals for heart rate variability analysis is considered important, and further investigation of this area is encouraged.
Keywords: Analysis of Variance, Electric Power Supplies, Electrocardiography - statistics & numerical data, Heart Rate, Signal Processing, Computer-Assisted, Analysis of Variance, Electric Power Supplies, Electrocardiography - statistics & numerical data, Heart Rate, Signal Processing, Computer-Assisted
Editorial
01 April 2024 : Editorial
Editorial: Forty Years of Waiting for Prevention and Cure of HIV Infection – Ongoing Challenges and Hopes for Vaccine Development and Overcoming Antiretroviral Drug ResistanceDOI: 10.12659/MSM.944600
Med Sci Monit 2024; 30:e944600
In Press
06 Mar 2024 : Clinical Research
Prevalence and Variation of Medical Comorbidities in Oral Surgery Patients: A Retrospective Study at Jazan ...Med Sci Monit In Press; DOI: 10.12659/MSM.943884
08 Mar 2024 : Clinical Research
Evaluation of Foot Structure in Preschool Children Based on Body MassMed Sci Monit In Press; DOI: 10.12659/MSM.943765
15 Apr 2024 : Laboratory Research
The Role of Copper-Induced M2 Macrophage Polarization in Protecting Cartilage Matrix in OsteoarthritisMed Sci Monit In Press; DOI: 10.12659/MSM.943738
07 Mar 2024 : Clinical Research
Knowledge of and Attitudes Toward Clinical Trials: A Questionnaire-Based Study of 179 Male Third- and Fourt...Med Sci Monit In Press; DOI: 10.12659/MSM.943468
Most Viewed Current Articles
17 Jan 2024 : Review article
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
16 May 2023 : Clinical Research
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
01 Jan 2022 : Editorial
Editorial: Current Status of Oral Antiviral Drug Treatments for SARS-CoV-2 Infection in Non-Hospitalized Pa...DOI :10.12659/MSM.935952
Med Sci Monit 2022; 28:e935952