28 January 2022>: Animal Study
Feasibility of Using a Portable MyotonPRO Device to Quantify the Elastic Properties of Skeletal Muscle
Ya-Peng Li 1ABCE , Chun-Long Liu 2ACDF* , Zhi-Jie Zhang 1AEF*DOI: 10.12659/MSM.934121
Med Sci Monit 2022; 28:e934121
Table 3 The inter-operator reliability of the MyotonPRO in measurement of BBM stiffness.
Operator 1 (N/m) | Operator 2 (N/m) | SEM (N/m) | ICC | 95% CI | MDC (N/m) | |
---|---|---|---|---|---|---|
Dominant arm | 186.21±17.09 | 182.39±15.94 | 4.81 | 0.86 | 0.49–0.96 | 13.33 |
SEM – standard error of the mean; ICC – intraclass correlation coefficient; MDC – minimum detectable change, 95% CI – 95% confidence interval. |