Logo Medical Science Monitor

Call: +1.631.470.9640
Mon - Fri 10:00 am - 02:00 pm EST

Contact Us

Logo Medical Science Monitor Logo Medical Science Monitor Logo Medical Science Monitor

24 June 2021: Clinical Research

An Individualized Contrast-Enhanced Liver Computed Tomography Imaging Protocol Based on Body Mass Index in 126 Patients Seen for Liver Cirrhosis

Jian Jiang ABCEG , Maowei Zhang BD , Yuan Ji BCDF , Chunfeng Li CD , Xin Fang BF , Shuyuan Zhang B , Wei Wang CF , Lijun Wang EF , Ailian Liu ACEF*

DOI: 10.12659/MSM.932109

Med Sci Monit 2021; 27:e932109

Table 1 CT acquisition, contrast injection protocols, and reconstruction parameters.

ParameterGroup A (80 kV)Group B (100 kV)Group C (120 kV)
BMI (kg/m)≤2424.1–28≥28.1
CT scanning
 Tube voltage (kV)80100120
 Tube current (mA)AutoAutoAuto
 FOV (mm)363636
 Gantry rotation time (sec)0.80.80.8
 Helical pitch0.984: 10.984: 10.984: 1
Reconstruction
 Reconstructed section thickness (mm)5.005.005.00
 Reconstructed section increment (mm)5.005.005.00
 Reconstruction algorithm40% ASIR40% ASIR40% ASIR
 Reconstruction kernelSoft tissueSoft tissueSoft tissue
Contrast Injection
 Contrast mediumUltravist-300Ultravist-300Ultravist-300
 Iodine amount (mg I/kg)352440550
 Saline (ml)303030
 Injection duration (sec)253030
 Bolus speed (ml/sec)
BMI – body mass index; ASIR – adaptive statistical iterative reconstruction; – body weight; ASIR – adaptive statistical iterative reconstruction; FOV – field of view.

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750